Hipot Tester UX36-ADC-0507-0100-TPT
Price on request
Details
+49 421 277 9999
The UX36-0507-0100-TPT hipot tester is a version of the UX-series test sets with 5kV AC and 7kV DC. The units of the UX series are available in an elegant industrial housing including high-voltage equipment, as well as test adapter, 2-hand operation and safety precautions.
Highlights
- 5000V AC and 7000V DC (selectable)
- Ramping
- 4-wire-contact monitoring
- Wire break monitoring
- Launch control
- Programmable test parameters
- RS232, digital I/O (SPS), CAN
- Software, DLL, LabView, Active-X (option)
Test methods
High voltage test AC
- AC voltage: 100 V - 5000 V
- AC current: 0.5 - 100 mA
- Short circuit current: 200 mA
- Transformer capacity: 500 VA
- Output frequency 50 Hz / 60 Hz
High voltage test DC
- DC voltage: 100 V - 7000 V
- DC current: 0.1 - 70 mA
- Short circuit current: 200 mA
- Transformer power: 500VA
Technical data
- Desktop unit with dry-type transformer
- Potential free (suitable for use with test probes)
- Regulated test voltage
- Burning function in AC mode
- High voltage output on the rear side
- Error message: acoustic, optical, interface
- Zero voltage switch(to conserve test object)
- ACTUAL- and SET POINT-display (current/voltage)
- Max. 1 sec. Burning function (AC)
- Keyboard lock
- Acoustic signaling adjustable
- Display brightness control
- Start mode selectable
- Data management program with your PC (optional)
- Ramp times and ramp types
- Verifying time scaling
- Test time: 1-99s / 1-99 minutes and infinitely
- Minimum current monitoring
- Integrated program memory
- Large LED display and various control buttons
- CAN bus for expansions
- Digital I/O:
- START / STOP
- PASS / FAIL
- Test in progress
- Terminals:
- 2 HV sockets potential free
- Warning lamp
- Foot switch, test-cage, light curtain, prod.-line, two-hand-operation
- Highest safety standard:
- Key switch preventing unauthorized use
- High voltage absolutely potential-free
- Two fold isolated transformer and measurement converter
- Status indication
- Socket for external pilot lamp
- Integrated warning buzzer
- Insulated sockets
- Security activation knob in compliance with VDE0104
- Safety Circuit with forcibly actuated HV relays
- Power supply: 230V, 50/60 Hz, 8A max.
- Operating temperature: +5° to +45°
- Humidity: 70% (non condensing)
- Height: 2000 m above NN
- Compact industrial housing (RAL 7035)
- Dimensions (W/H/D): 590/210/520 mm
- Weight: 20 kg
- CE-conformity, according to EN 50 191
What's in the box
- Interlock connector
- Key set
- Power cable
- Operation manual
- Factory calibration
Options
- HTP06N High voltage test pistols
- HVP06C HV cable with plug
- KLD-SW alligator clips black
- WK36 Warning lights set
Related products
- Devices: hipot testers 36 series
- Signal lamp: red / green
- Conforming to all standards: VDE0104, EN50191
- Status:
- READY FOR OPERATION
- READY TO SWITCH ON
- IN OPERATION
- Voltage: 230V
- Cable length: 2m
- Side mounting for test waggon or table
- Ergonomically placed
- HV probe type: HTP
- Side: right
- Colour: white
- HV Tester: UB36, UG36, UH36, UX36
- With foot protection cover
- Start via control-IO
- With plug: SUB-D 9-pole
- Access to control interface
- Cable length: 2m
- Rectangle design
- "Danger High Voltage"
- Material: plactic
- Size: 74mm x 210mm
- Colour: yellow/black
- EN50191 / DIN 40008
- Robust 2-wire matrix
- Measuring points freely connectable
- Test connection via Harting plug
- Continuity tests
- High voltage tests
- Insulation resistance tests
- Surge voltage tests
- Reliable HV Probe without plug
- Highly flexible HV connecting cable
- Operating voltage 10 kVDC / 7 kVAC
- Hard-wearing tungsten test prod
- CE conformity, according EN 50 191
- Standard-compliant, multi-vendor calibration based on ISO 17025
- Hipot and Insulationtester
- Traceable to DKD
- Cleaning and adjustment
- Repair service
- HV-Cable, two-pole
- Length 8 m
- HV-right-angle plug HVP06CW
- Opposite plug: lamellar plug
- 7 kV AC / 10 kV DC, max. 10 A
- Reliable HV Probe
- Highly flexible HV connecting cable
- Operating voltage 10 kV DC / 7 kVAC
- Hard-wearing tungsten test probe
- CE conformity, according EN 50 191
- HV-Cable, two-pole
- Length 2 m
- HV-plug HVP06C
- 7 kV AC / 10 kV DC, max. 10 A
- HV-Cable, two-pole
- Length 2 m
- HV-plug HVP06C
- 7 kV AC / 10 kV DC, max. 10 A
- All-Round-Test-Cage to connect
with safety testers and hipot testers - Dimension (W/H/D):
Outside: 450 / 515 / 495 mm
Inside: 360 / 350 / 450 mm - Viewable transparent swinging hood
- Damping gas spring
- Safety limit switch
- CE conform with EN 50191
- Devices: series 36 & 400
- Complete library
- License for 1 test system
- Including CD with drivers and manual
- Conforming to all standards
- Emergency stop for tester
- Load: 16 A
- Cable length: 5 m
- Including cable set for safety tester ST400 series
- Dimension (W/H/D):
Outside: 450 / 515 / 495 mm
Inside: 360 / 350 / 450 mm - Completely mounted and wired for plug&play
- PVC design-housing
- Viewable transparent swinging hood
- Damping gas spring
- Safety limit switch
- CE conform with EN 50191
- Rectangle design
- "No unauthorised entry"
- Material: plactic
- Size: 200mm x 300mm
- Colour: red/white
- Log printer in own plastic table casing
- Not to be built-in.
- Baud rate 9600
- Portable posts chain
- Convex base plate
- Colour: Yellow-Black
- Two-parts
- Height: 1100mm / D=32cm
- With safety socket
- For easy testing of test objects
- L and N bridged/connected
- In plastic casing
- Including cable-set for hipot tester:
36 series (UH36/UX36) - Ergonomically design
- Push-start-buttons: 2
- Normed distance between buttons
- Connecting cable: 2.5 m
- Side mounting for test waggon or table
- Ergonomically placed
- HV probe type: HTP
- Side: left
- Colour: white
- HV Tester: UB36, UG36, UH36, UX36
- Start via control-IO
- With plug: SUB-D 9-pole
- Access to control interface
- Cable length: 2m
- All-Round-Test-Cage to connect
with safety testers and hipot testers - Dimension (W/H/D):
Outside: 670 / 515 / 495 mm
Inside: 565 / 350 / 450 mm - Viewable transparent swinging hood
- Damping gas spring
- Safety limit switch
- CE conform with EN 50191
- Devices: series 36 & 400
- Complete library
- License for 1 test system
- Including CD with drivers and manual
- HV-Cable, two-pole
- Length 3 m
- HV-plug HVP06C
- With contact terminal
- 7 kV AC / 10 kV DC, max. 10 A
- Reliable HV Probe
- Highly flexible HV connecting cable
- Operating voltage 10 kVDC / 7 kVAC
- Hard-wearing tungsten test prod
- CE conformity, according EN 50 191
- Text and screen displayed steps
- Single step-mode
- Large PASS-FAIL display
- Network integration
- Data-format: xml / Access / SQL
- Statistic-functions
- Barcorde function
- Devices: High Voltage Tester 36-Series
- Remote control via:Ethernet / USB / RS232
- Integration label-printer
- HV-Cable, two-pole
- Length 2 m
- HV-plug HVP06C
- 2 x counter plug
- 7 kV AC / 10 kV DC, max. 10 A
- Including cable set for safety tester ST400 series
- Dimension (W/H/D):
Outside: 670 / 515 / 495 mm
Inside: 565 / 350 / 450 mm - Viewable transparent swinging hood
- Damping gas spring incl. final pos. damping
- Safety limit switch
- CE conform with EN 50191
- Triangle design
- With flash arrow
- Material: plactic
- Size: 200mm x 200mm
- Colour: yellow/black
- DIN 40008
- Devices: 28, 36 & 400
- Wide grip: brass
- Toothed jaws
- Bonding: 30 mm
- For laminated plug
- With safety socket and test nipple
- For easy testing of test objects
- In plastic casing
- Devices: 36 series
- Mounted on test trolley
- Total hight of lamps 1850mm
- Standard: EN 50191, VDE0104
- Lights: Red/Green
- Function: operation, switch on, ready
- Cable length: 2m