Hipot Tester UG36
€3,240.00*
Details
+49 421 277 9999
UG36 is a digital 6kVDC hipot tester with RS232, digital I/O control interface and CAN. The UG36 hipot tester can be used in production lines as well as in workshops and laboratories. Special function is the contact monitoring.
Highlights
- 6kV DC hipot tester UG36
- Reliable burn function
- 4-wire-contact-monitoring
- Contactor breakage
- Automatically START
- Test parameters programmable
- Potential free in accordance to EN50191
- RS232, digital I/O, CAN
- Release current adjustable
- Terminals (PLC, footswitch, cave, 2-hand-operation)
Test methods
- DC voltage: 100V - 6000V
- Resolution: 10V
- Accuracy: 1% ± 2 digits
- DC current: 0.1mA - 10mA
- Resolution: 0.1mA
- Accuracy: 2% ± 3 digits
- Release current programmable
- Switchable tests:
- "test"
- "burn"
- Burning out of the faulty positions
- Output voltage regulated
Technical data
- Error message: optic, acoustic, interface
- Zero voltage switch (protects your test object)
- Target and true values displayed permanently
- Key lock function
- Buzzer options
- LED display brightness selectable
- Printer-function: OK, NOK, always
- Ramp time and ramp type
- Test duration calibration
- Test time: 1-99s / 1-99min and permanently
- Minimal current monitoring
- Memory for 9 parameter settings
- LED's (operation, HV, remote, burn)
- 2 buttons (operation, HV)
- CAN-interface to extend in testing system
- Digital I/O:
- START / STOP
- PASS / FAIL
- Test in progress - Digital I/O:
- START / STOP
- PASS / FAIL
- Test in progress
- Terminals:
- 2 HV sockets potential free
- Warning lamp
- Foot switch, test-cage, light curtain, prod.-line, two-hand-operation
- Highest safety standard:
- Key switch preventing unauthorized use
- High voltage absolutely potential-free
- Two fold isolated transformer and measurement converter
- Status indication
- Socket for external pilot lamp
- Integrated warning buzzer
- Insulated sockets
- Security activation knob in compliance with VDE0104
- Safety Circuit with forcibly actuated HV relays
- Supply: 230V, 50/60 Hz, 8A max.
- Operating temperature: +5° up to +45°
- Air humidity: 70% (non-condensing)
- Attitude: 2000m
- Compact industrial casing (W/H/D): 307/164/273 mm
- Weight: 15.8 kg
- CE-compliant, in accordance with EN 50 191
What's in the box:
- Safety circuit plug
- Power cable
- Operation manual
- Factory calibration
Options
- HTP06N HV probe
- HVP06C HV cable with plug
- KLD-SW dolphin clamp black
- WK36 warning lamp
- RC carrying handle (holder)
- HVS06N HV sockets backside
Related products
- Devices: hipot testers 36 series
- Signal lamp: red / green
- Conforming to all standards: VDE0104, EN50191
- Status:
- READY FOR OPERATION
- READY TO SWITCH ON
- IN OPERATION
- Voltage: 230V
- Cable length: 2m
- Triangle design
- With flash arrow
- Material: plactic
- Size: 200mm x 200mm
- Colour: yellow/black
- DIN 40008
- Devices: series 36 & 400
- Complete library
- License for 1 test system
- Including CD with drivers and manual
- With safety socket and test nipple
- For easy testing of test objects
- In plastic casing
- Including cable set for safety tester ST400 series
- Dimension (W/H/D):
Outside: 670 / 515 / 495 mm
Inside: 565 / 350 / 450 mm - Viewable transparent swinging hood
- Damping gas spring incl. final pos. damping
- Safety limit switch
- CE conform with EN 50191
- Side mounting for test waggon or table
- Ergonomically placed
- HV probe type: HTP
- Side: right
- Colour: white
- Devices: 28, 36 & 400
- Wide grip: brass
- Toothed jaws
- Bonding: 30 mm
- For laminated plug
- Rectangle design
- "Danger High Voltage"
- Material: plactic
- Size: 74mm x 210mm
- Colour: yellow/black
- EN50191 / DIN 40008
- Reliable HV Probe
- Highly flexible HV connecting cable
- Operating voltage 10 kVDC / 7 kVAC
- Hard-wearing tungsten test prod
- CE conformity, according EN 50 191
- Including cable-set for hipot tester:
36 series (UH36/UX36) - Ergonomically design
- Push-start-buttons: 2
- Normed distance between buttons
- Connecting cable: 2.5 m
- HV Tester: UB36, UG36, UH36, UX36
- Start via control-IO
- With plug: SUB-D 9-pole
- Access to control interface
- Cable length: 2m
- All-Round-Test-Cage to connect
with safety testers and hipot testers - Dimension (W/H/D):
Outside: 450 / 515 / 495 mm
Inside: 360 / 350 / 450 mm - Viewable transparent swinging hood
- Damping gas spring
- Safety limit switch
- CE conform with EN 50191
- Conforming to all standards
- Emergency stop for tester
- Load: 16 A
- Cable length: 5 m
- Robust 2-wire matrix
- Measuring points freely connectable
- Test connection via Harting plug
- Continuity tests
- High voltage tests
- Insulation resistance tests
- Surge voltage tests
- Rectangle design
- "No unauthorised entry"
- Material: plactic
- Size: 200mm x 300mm
- Colour: red/white
- Text and screen displayed steps
- Single step-mode
- Large PASS-FAIL display
- Network integration
- Data-format: xml / Access / SQL
- Statistic-functions
- Barcorde function
- Devices: High Voltage Tester 36-Series
- Remote control via:Ethernet / USB / RS232
- Integration label-printer
- Portable posts chain
- Convex base plate
- Colour: Yellow-Black
- Two-parts
- Height: 1100mm / D=32cm
- Devices: series 36 & 400
- Complete library
- License for 1 test system
- Including CD with drivers and manual
- With safety socket
- For easy testing of test objects
- L and N bridged/connected
- In plastic casing
- Including cable set for safety tester ST400 series
- Dimension (W/H/D):
Outside: 450 / 515 / 495 mm
Inside: 360 / 350 / 450 mm - Completely mounted and wired for plug&play
- PVC design-housing
- Viewable transparent swinging hood
- Damping gas spring
- Safety limit switch
- CE conform with EN 50191
- Side mounting for test waggon or table
- Ergonomically placed
- HV probe type: HTP
- Side: left
- Colour: white
- Log printer in own plastic table casing
- Not to be built-in.
- Baud rate 9600
- Reliable HV Probe without plug
- Highly flexible HV connecting cable
- Operating voltage 10 kVDC / 7 kVAC
- Hard-wearing tungsten test prod
- CE conformity, according EN 50 191
- Devices: 36 series
- Mounted on test trolley
- Total hight of lamps 1850mm
- Standard: EN 50191, VDE0104
- Lights: Red/Green
- Function: operation, switch on, ready
- Cable length: 2m
- Reliable HV Probe
- Highly flexible HV connecting cable
- Operating voltage 10 kV DC / 7 kVAC
- Hard-wearing tungsten test probe
- CE conformity, according EN 50 191
- HV Tester: UB36, UG36, UH36, UX36
- With foot protection cover
- Start via control-IO
- With plug: SUB-D 9-pole
- Access to control interface
- Cable length: 2m
- Standard-compliant, multi-vendor calibration based on ISO 17025
- Calibration of Hipot Tester / HV tester
- Traceable to DKD
- Cleaning and adjustment
- Repair service
- HV-Cable, two-pole
- Length 2 m
- HV-plug HVP06C
- 7 kV AC / 10 kV DC, max. 10 A